Ozone analyzer
Ozone analyzer that uses UV photometry technology
Uses unique UV Photometry technology with a patented LED: the O342M is the first commercially certified ozone analyzer not using a mercury lamp as a source of excitation.
Advantages
- User programmable ranges and average times
- Real time calibration graph
- Full remote emulation of the analyzer
- Graphic Liquid Crystal (LCD) display
The outcome is an ultra compact and lightweight (rack 3U), easy-to-use analyzer capable of measuring ozone at ppb levels.
Applied to ozone measurement, the universally known UV absorption principle consists in measuring UV absorption of ozone molecules. Ozone concentration is determined by difference between UV absorption of the gas sample and the sample without ozone after filtration performed by a catalytic converter.
Discover also our instruments for the measurement and analysis of suspended particles : PM10, PM2.5 or PM1.
- Span drift : less than 0.5 % / 24 h less than 1 %
- Lower detectable limit : 0.4 ppb
- Response time : automatic and programmable (minimum 20 sec)
- zero drift : less than 0.5 ppb / 24 h less than 1 ppb / 7 days
- Continuous measurement of ozone (O3) in ambient air
- Continuous measurement in indoor air
- Impact studies on the effects of ozone